Invacare Softform PREMIER ORIGINAL, Softform A&E TROLLEY MATTRESS, Softform BARIATRIC MATTRESS, Softform EXCEL, Softform MAXIGLIDE, Softform PREMIER, Softform PREMIER VISCO Manual do usuário

  • Olá! Sou um chatbot de IA treinado para ajudá-lo com o Invacare Softform PREMIER ORIGINAL Manual do usuário. Já revisei o documento e posso oferecer explicações claras e úteis.
Invacare® Softform®
en Mattresses
UserManual..................................3
deMatratzen
Gebrauchsanweisung............................17
frMatelas
Manueld'utilisation.............................33
esColchones
Manualdelusuario.............................49
itMaterassi
Manualed’uso.................................65
nlMatrassen
Gebruiksaanwijzing..............................81
ptColchões
Manualdeutilização............................97
svMadrasser
Bruksanvisning.................................113
noMadrasser
Bruksanvisning.................................127
daMadrasser
Brugsanvisning.................................141
ThismanualMUSTbegiventotheuseroftheproduct.
BEFOREusingthisproduct,readthismanualandsaveforfuturereference.
©2017InvacareCorporation
Allrightsreserved.Republication,duplicationormodicationinwholeorinpartisprohibited
withoutpriorwrittenpermissionfromInvacare.Trademarksareidentiedby™and®.Alltrademarks
areownedbyorlicensedtoInvacareCorporationoritssubsidiariesunlessotherwisenoted.
Contents
ThismanualMUSTbegiventotheuseroftheproduct.
BEFOREusingthisproduct,readthismanualandsaveforfuture
reference.
1General.........................................4
1.1Generalinformation............................4
1.2Symbolsinthisusermanual......................4
1.3Compliance...................................4
1.4Warranty.....................................5
1.5Limitationofliability............................5
1.6Intendeduse..................................5
1.7Servicelife...................................5
2Safety..........................................6
2.1Safetyinformation..............................6
2.2Symbolsontheproduct.........................7
3Components......................................8
3.1Productdescription.............................8
3.2Components..................................8
4Usage..........................................9
4.1Safetyinformation..............................9
5Transport........................................10
5.1Safetyinformation..............................10
6Maintenance.....................................11
6.1Inspection....................................11
6.2Cleaningandcare..............................11
7AfterUse........................................13
7.1Storage......................................13
7.2Re-Use......................................13
7.3Disposal.....................................13
8TechnicalData....................................14
8.1GeneralData..................................14
8.2Materials....................................15
8.3EnvironmentalParameters........................15
InvacarSoftform®
1General
1.1Generalinformation
Essentialnursingcareispivotalinpressureulcerprevention.
Thesemattresseswillpositivelycontributetotheoutcome
ofapressureulcerpreventioncareplan.
Education,clinicaljudgementandactionbasedplanning
basedonvulnerabilityarefundamentalfactorsinthe
preventionofpressureulcers.
Arangeofassessmentscalescanbeusedasaformalmethod
ofassessingriskfrompressureulcerdevelopment,and
shouldbeusedinconjunctionwithaninformalassessment
(informednursingjudgement).Informalassessmentis
consideredtobeofgreaterimportanceandclinicalvalue.
Thisusermanualcontainsimportantinformationaboutthe
handlingoftheproduct.Inordertoensuresafetywhen
usingtheproduct,readtheusermanualcarefullyandfollow
thesafetyinstructions.
ForfurtherinformationcontactInvacareinyourcountry
(addressesseebackpageofthisusermanual).
Toaccessvaluableinformationandusefullinksfor
PressureAreaCaretrainingandeducation,referto
www.thinkpressurecare.co.uk.
1.2Symbolsinthisusermanual
InthisUserManualwarningsareindicatedbysymbols.
Thewarningsymbolsareaccompaniedbyaheadingthat
indicatestheseverityofthedanger.
WARNING
Indicatesapotentiallyhazardoussituation
whichifnotavoidedcouldresultindeathor
seriousinjury.
CAUTION
Indicatesapotentiallyhazardoussituation
whichifnotavoidedcouldresultinproduct
damage,minorinjuryorboth.
IMPORTANT
Indicatesahazardoussituationwhichifnot
avoidedcouldresultindamagetotheproduct.
Givesusefultips,recommendationsand
informationforefcient,trouble-freeuse.
Thisproductcomplieswiththedirective
93/42/EECformedicalproducts.Thelaunch
dateforthisproductisspeciedintheCE
declarationofconformity.
Manufacturer
1.3Compliance
Qualityisfundamentaltothecompany’soperation,working
withinthedisciplineofISO13485.
AllInvacareSoftformMattressproductsfeaturetheCEmark,
incompliancewiththeMedicalDeviceDirective93/42/EEC
Class1.
Invacareiscontinuouslyworkingtowardsensuringthatthe
company’simpactontheenvironment,locallyandglobally,
4
PALIN-010-C
General
isreducedtoaminimum.WeonlyuseREACHcompliant
materialsandcomponents.
ForfurtherinformationpleasecontactInvacareinyour
country(addressesseebackpageofthismanual).
1.4Warranty
Weprovideamanufacturerswarrantyfortheproduct
inaccordancewithourGeneralTermsandConditionsof
Businessintherespectivecountries.Guaranteeclaims
canonlybemadethroughtheproviderfromwhomthe
appliancewasobtained.
1.5Limitationofliability
Invacareacceptsnoliabilityfordamagearisingfrom:
Non-compliancewiththeusermanual
Incorrectuse
Naturalwearandtear
Incorrectassemblyorset-upbythepurchaserora
thirdparty
Technicalmodications
Unauthorizedmodicationsand/oruseofunsuitable
spareparts
1.6Intendeduse
Thispressureredistributionmattressisintendedtobeused
inconjunctionwithanappropriatelysizedbedframe,as
partofanoverallpressureulcerpreventionprogramofcare.
Anyotheruseisprohibited.
Thisproducthasbeendesignedtodelivereffectivepressure
redistributiontousers,whentheproductisinnormaluse
whichisdenedbyInvacareaswhenthesupportsurface
iscoveredwithacotton,cottoncombinationorlinen
bedsheet,andanyoneofthesewouldbetheonlyitem
deployedbetweenthesupportsurfaceandtheuser.
1.7Servicelife
Weestimatealifeexpectancyofveyearsforthese
products,providedtheyareusedinstrictaccordance
withtheintendeduseassetoutinthisdocumentand
allmaintenanceandservicerequirementsaremet.The
estimatedlifeexpectancycanbeexceedediftheproduct
iscarefullyusedandproperlymaintained,andprovided
technicalandscienticadvancesdonotresultintechnical
limitations.Thelifeexpectancycanalsobeconsiderably
reducedbyextremeorincorrectusage.
Thefactthatweestimatealifeexpectancyforthese
productsdoesnotconstituteanadditionalwarranty.
PALIN-010-C5
InvacarSoftform®
2Safety
2.1Safetyinformation
WARNING!
Donotusethisproductoranyavailableoptional
equipmentwithoutrstcompletelyreadingand
understandingtheusermanualsupplied.Invacare
productmanualsareavailableonyourlocal
Invacarewebsiteoratyourlocaldealer.Ifyou
areunabletounderstandthewarnings,cautions
orinstructionspleasecontactahealthcare
professional,dealerortechnicalpersonnelbefore
attemptingtousethisequipmentotherwise,
injuryordamagemayoccur.
WARNING!
Riskofdevelopingpressureulcers
Bedsheetsmustbelooselytted,withcreases
smoothedout.Caremustalwaysbetakento
ensurethatthesupportsurfaceincontactwith
theuseriskeptfreefromcrumbsandother
fooddebris,andthatdripcables,stents,and
otherforeignobjectsdonotbecomeentrapped
betweentheuserandthepressurereducing
surfaceofthemattress,asthismayresultinthe
developmentofpressureulcers.
WARNING!
Invacareproductsarespecicallydesigned
andmanufacturedforuseinconjunctionwith
Invacareaccessories.Accessoriesdesignedby
othermanufacturershavenotbeentestedby
Invacareandarenotrecommendedforusewith
Invacareproducts.
Theintroductionofcertainthirdpartyproducts
betweenthemattresssurfaceandtheusermay
reduceorimpedetheclinicaleffectivenessofthis
product.
’Thirdpartyproducts’mayinclude,butare
notlimitedtoitemsincludingunderblankets,
plasticsheetsandsheepskins,etc.Heatedover
blanketsmustonlybeusedinconsultationwitha
suitablyqualiedhealthcareprofessional,asan
increaseintemperaturecanincreasetheriskof
developingpressureulcers.
WARNING!
Riskofreorexplosion!
Acigarettecanburnaholeinthebedsurface
andcausedamagetothemattress.Also,patient
clothing,bedsheets,etc,maybecombustible
andcauseare.Failuretoobservethiswarning
canresultinaseverere,propertydamageand
causephysicalinjuryordeath.
Donotuseinoxygenrichenvironments.
Donotsmoke.
6PALIN-010-C
Safety
IMPORTANT!
Theinformationcontainedinthisdocumentis
subjecttochangewithoutnotice.
Checkallpartsforshippingdamageandtest
beforeusing.
Incaseofdamage,donotuse.
ContactInvacareforfurtherguidance/informa-
tion.
2.2Symbolsontheproduct
Donotpierceor
cut
Linedry
DonotdrycleanCEconform
xxx kg
Userweightlimit*
80°
Recommended
80°C
Donotputnear
ame
Donotiron
DonotbleachTumbledrylow
heat
Dateof
manufacture
*Maximumuserweightaspersection8TechnicalData,
page14.
PALIN-010-C
7
InvacarSoftform®
3Components
3.1Productdescription
TheInvacarSoftform®mattressoffersexceptionalpressure
reducingqualities,specictotheindividualsneeds.By
promotingevenweightdistribution,theyreducetissue
pressureatpotentiallyvulnerableareas.
Thewater-resistantcoverprovidesavapour-permeable,
multistretchsurface,topromotepatientcomfortandto
maximizetheeffectivenessofthefoamcore.
3.2Components
Thefollowingcomponentsareincludedwithinthescope
ofdelivery*:
A
B
C
D
E
A
Multi-stretchwaterproofpolyurethanecover
B
Castellatedfoaminsert
C
Singlepiecefoam“U”core
D
Toughedpolyurethanecoatedbase
E
UserManual
*ThisexampleisbasedontheSoftform®Premier,other
mattresscongurationsmayvary.
8PALIN-010-C
Usage
4Usage
4.1Safetyinformation
1.Removeallpackagingbeforeuse.
2.Placethemattressdirectlyontheframeofthebed.
Themattressisdesignedforbedswithadjustablelying
surface.
WARNING!
Itisveryimportantforthepatienttoreposition
themselves,ortoberepositioned,onaregular
basis.Thismustbebasedontheclinical
judgementofaqualiedhealthcareprofessional.
Thisrelievespressurewhichhelpspreventboth
tissuecompressionandpotentialulcerformation.
Alwaysconsultaqualiedhealthcare
professionalbeforeusingtheproduct.
Monitorthepatientfrequently.
CAUTION!
Makesurethattheprintedsideofthemattress
coveralwaysfacesupwards.
Ensurethatthedistancebetweenthesurface
ofthemattressandthetopofthesiderailis
atleast220mm.Ifthisisnotachievablearisk
assessmentmustbecarriedout.
IMPORTANT!
Ifholesarepresentinthemattresscover,
thereisadangerthatliquidsmayingressand
contaminationmayoccur.
Medicalequipmentincludinginfusionpumps
andmonitorsshouldbeattachedtoappropriate
bedaccessories.
Forhomeusecommoncausesofdamage
includecigaretteburnsandtheclawsofpets
thatpuncturesheets,allowinguidingressand
staining.
IMPORTANT!
Riskofdamagetothemattresscover
Topreventaccidentaldamage,donotplace
hypodermicneedles,venons,scalpelsorother
similarlysharpobjectsontothemattress.
Ensurethatallvenonsaretapeddown
correctlywithnosharpedgesexposed.
Whenusingpatienttransferaids,careshould
betakennottodamagethemattress.All
transferaidsshouldbecheckedforanysharp
edgesorburrsbeforeuseasthesecandamage
themattress.
Makesurethatthemattressesarenotjammed
ordamagedbysharpedgeswhenusedon
bedswithanadjustableframe.
Whenusingthemattressonaprolingbed
ensurethatthekneebreakisusedbeforethe
backrest.
PALIN-010-C9
InvacarSoftform®
5Transport
5.1Safetyinformation
IMPORTANT!
Takecarewhenhandlingmattressestoensure
nodamagetothecover.Itisrecommended
thattwopeoplelift/carrymattresses.
Avoidcontactwithjewellery,nails,abrasive
surfacesetc.
Donotdragmattresses.
Avoidcontactwithwall,doorframes,door
catchesorlocksetc.
Donottransportinrollcagesunlesscompletely
protectedfromthesharpedgesofthecage.
1.Refertothestorageandshippingconditionsinsection
8.3EnvironmentalParameters,page15.
10PALIN-010-C
Maintenance
6Maintenance
6.1Inspection
Checkmattresses(foamandcover)forstrike-through(this
mayincludeuidingress,stains,ripsordamage)afterthe
releaseofeachpatient,afterendingoftheperiodofuseor
onaminimummonthlybasis(dependingonwhichoccurs
rst)byasuitablyqualiedandcompetentperson.
Checkmattress
1.Unzipthecovercompletely.
2.Checkforanystainingonthewhiteundersideofthe
cover.
3.Checkforanystainingontheinteriorfoam.
4.Replaceanystaineditemsanddisposeofasperlocal
authorityprocedure.
6.2Cleaningandcare
IMPORTANT!
Allcleaningagentsanddisinfectantsusedmust
beeffective,compatiblewithoneanotherand
mustprotectthematerialstheyareusedtoclean.
Forfurtherinformationondecontamination
inhealthcareenvironments,pleaserefer
toTheNationalInstituteforClinical
Excellence’guidelinesonInfectionControl
www.nice.org.uk/CG139andyourlocalcountry
infectioncontrolpolicy.
Cleaningcovers
(Removalofcontaminantssuchasdustandorganicmatter)
1.Removeallcoversforlaundering.
2.Launderthecoverswiththerecommendedtemperature
at
80°Cusingadiluteddetergentsolution(Instructions
onlabel).
IMPORTANT!
Washingathighertemperatureswillcause
shrinkage.
Dryingcovers
1.Hangmattresscoversfromalineorbaranddripdryin
acleanindoorenvironment.
or
Tumbledryonalowheatsetting.
IMPORTANT!
Tumbledrysettingmustnotexceed40°C.
Donottumbledryforlongerthan10minutes.
Drythoroughlybeforere-ttingtothefoam.
Disinfectingcovers
(Reducingthenumberofmicroorganisms)
Pleasecontactyourhygienespecialistintheeventof
contamination.
IMPORTANT!
Ensurethatanyresidualdetergenthasbeen
removedwithcleanwaterpriortodisinfection.
Lightsoilage
PALIN-010-C
11
InvacarSoftform®
1.Wipedownthecoverwitha0.1%ChlorineSolution
(1,000ppm).
2.Rinsethecoverwithcleanwaterusingasingleuse
nonabrasivecloth.
3.Drythecoverthoroughly.
Heavysoilage
Wherethemattressisbadlysoiled,werecommendcleaning
withadilutecleaningsolutionat80°Cinthewashing
machine.
Largespillagesofbloodshouldbeabsorbedand
removedwithpapertowelsrst,followedbyas
above.
1.Cleanupallspillagesofbodilyuidsi.e.blood,urine,
faeces,sputum,woundexudaterandallotherbodily
secretionsassoonaspossibleusinga1%Chlorine
Solution(10,000ppm).
2.Rinsewithcleanwaterusingasingleusenonabrasive
cloth.
3.Drythecoverthoroughly.
IMPORTANT!
Polyurethanecoatedfabricscanabsorbliquidsfor
shortperiodscausingatemporarychangetothe
polyurethanecharacteristics.Themattresscover
swellstemporarilyandismorevulnerableto
physicaldamageforaperiodafteritiscompletely
surfacedried,bywhichtimeitwillreverttoits
previousstate.
IMPORTANT!
1%ChlorineSolutionusedonaregularbasiscan
diminishthelifeofthecoverifnotrinsedand
driedproperly.
Donotusegranules.
WARNING!
Removecontaminatedfoamsfromuse.
CAUTION!
Keepclearofopenheatsources.
IMPORTANT!
Donotusephenols,alcohols,bleaches,or
otherabrasivematerials.
Replacingcovers
1.Unzipthecoverandremoveitcarefullyfromthefoam
core.
2.Placenewcoverontothefoamcore.
3.Thenclosethezipper.
IMPORTANT!
Ensurethatthecornersofthefoamcoreare
positionedcorrectlyintothecornersofthe
cover.
Ensurethatthecastillatedfoamisfacing
uppermostwhenpackedintoitscover.
12
PALIN-010-C
AfterUse
7AfterUse
7.1Storage
Toeasestorageandhandlingthemattressesaresometimes
compactrolleddependingoncustomerrequirements.This
innowayaffectstheperformanceandlongevityofthe
product.Whenthemattressisunrolledforthersttime,
allowittorecoverforaminimumof1hourbeforeuseto
optimizeperformance.
IMPORTANT!
Storemattressesinadryenvironment.
Storemattresseswithinaprotectivecover.
Storeitemsonclean,dry,off-ooringfreefrom
sharpedgestoavoidanypossibledamage.
Neverstoreotheritemsontopofamattress.
Donotstoremattressesnexttoradiatorsor
otherheatingdevices.
Protectmattressesfromdirectsunlight.
1.Refertothestorageandshippingconditionsinsection
8.3EnvironmentalParameters,page15.
7.2Re-Use
Acleaningrecordmustbekeptaspartofcleaningthe
system.
Theproductissuitableforrepeateduse.Thenumberof
timesitcanbeuseddependsonhowoftenandinwhich
waytheproductisused.
1.Beforereuse,cleantheproductthoroughly,
6.2Cleaningandcare,page11.
7.3Disposal
Thedisposalandrecyclingofuseddevicesandpackaging
mustcomplywiththeapplicablelocallegalregulations.
1.Ensurethatthemattressiscleanedpriortodisposalto
avoidanyriskofcontamination.
PALIN-010-C13
InvacarSoftform®
8TechnicalData
8.1GeneralData
Product
Firetesting
Graderef&ColourNominal
density
range
[kg/m3]
Nominal
hardness
range[N]
Maximumuser
weight[kg]
Weightof
product
[kg]
1)
PREMIER
EN597-1,2
BS7177:Crib5
RX39/120Pink
RX39/200Blue
3840
3840
105135
180200
247.614
MAXIGLIDE
EN597-1,2
BS7177:Crib5
RX39/120Pink
RX39/200Baltic
Blue
3840
3840
105135
180200
247.613.5
PREMIERVISCO
EN597-1,2
BS5852:Crib5
RX39/120Pink
RX39/200Blue
VC55085White
3840
3840
55
105135
180200
85
247.614
EXCEL
EN597-1,2
BS7177:Crib5
RX39/120Pink
RX39/200Blue
3840
3840
105135
180200
14010
PREMIERORIGINAL
EN597-1,2
BS7177:Crib5
RX39/120Orange
RX39/200Blue
3840
3840
105135
180200
247.613
BARIATRIC
MATTRESS
EN597-1,2
BS7177:Crib5
RX39/120Pink
RX39/200Blue
3840
3840
105135
180200
34015
A&ETROLLEY
MATTRESS
EN597-1,2
BS5852:Crib5
RX36/125Beige
RX39/200Blue
3537
3840
110140
180200
108
(TM20/TM40)
140(TM60)
5.5
1)
Basedontheweightofastandardsizemattress.Thiscanchangeifdifferentsizesareordered.
14
PALIN-010-C
TechnicalData
8.2Materials
Invacare®Softform®
Foam
PolyurethaneCombustionModiedHigh
ResilienceFoam
Cover
Polyurethanetransfercoatingonweft
knittedfabric
Cover(A&E
Trolley
mattress)
Polyurethanetransfercoatingonweft
knittedfabricwithenhancedelectrical
conductivity
Allmattresscomponentsarenotmadefromnaturalrubber
latex.
8.3EnvironmentalParameters
Operatingconditions
Ambienttemperature
10-35°C
Relativehumidity30%-75%Non-Condensing
Atmosphericpressure70-106kPa
Storageandshippingconditions
Ambienttemperature
-40-70°C
Relativehumidity
10%-100%
Atmosphericpressure50-106kPa
PALIN-010-C15
Notes
Inhaltsverzeichnis
DieseGebrauchsanweisungmussdemBenutzerdesProdukts
ausgehändigtwerden.LesenSiedieseGebrauchsanweisungvor
VerwendungdesProduktesundbewahrenSiesiefürdenweiteren
Gebrauchauf.
1Allgemein.......................................18
1.1AllgemeineInformationen........................18
1.2SymboleindieserGebrauchsanweisung..............18
1.3Konformität...................................19
1.4Garantie.....................................19
1.5BeschränkungderHaftung........................19
1.6Verwendungszweck.............................19
1.7Produktlebensdauer............................19
2Sicherheit.......................................21
2.1Sicherheitsinformationen.........................21
2.2SymboleamProdukt............................22
3AufbauundFunktion...............................23
3.1Produktbeschreibung............................23
3.2Komponenten.................................23
4Verwenden......................................24
4.1Sicherheitsinformationen.........................24
5Transport........................................26
5.1Sicherheitsinformationen.........................26
6Instandhaltung....................................27
6.1Inspektion....................................27
6.2ReinigungundPege............................27
7NachdemGebrauch...............................30
7.1Lagerung.....................................30
7.2Wiederverwendung.............................30
7.3Entsorgung...................................30
8TechnischeDaten..................................31
8.1AllgemeineDaten..............................31
8.2Materialien...................................32
8.3Umweltparameter..............................32
InvacarSoftform®
1Allgemein
1.1AllgemeineInformationen
EinehochwertigePegeistfürdieDekubitus-Prophylaxe
vonentscheidenderBedeutung.DieseMatratzenleisten
einenwichtigenBeitragzumErfolgeinesentsprechenden
Pegeplans.
Ausbildung,klinischesUrteilsvermögenund
MaßnahmenplanungentsprechendderAnfälligkeit
sindzentraleErfolgsfaktorenfürdieDekubitus-Prophylaxe.
AlsformaleMethodezurBeurteilungdesRisikosfür
dieEntwicklungeinesDekubituskanneineReihevon
Beurteilungsskalendienen,dieimKontexteinerinformellen
Beurteilung(BeurteilungdurchqualiziertesPegepersonal)
angewandtwerdenmüssen.DieinformelleBeurteilungwird
alswichtigerundvonhöheremklinischenWerterachtet.
DieseGebrauchsanweisungenthältwichtigeInformationen
zurHandhabungdesProdukts.LesenSiedie
Gebrauchsanweisungsorgfältigdurch,undbefolgenSiedie
Sicherheitsanweisungen,damiteinesichereVerwendungdes
Produktsgewährleistetist.
WeitereInformationenerhaltenSiebeiderNiederlassung
vonInvacareinIhremLand(AnschriftenndenSieaufder
RückseitedieserGebrauchsanweisung).
WertvolleInformationenundhilfreicheLinkszurAus-und
WeiterbildungimBereichDekubitus-PegendenSieunter
www.thinkpressurecare.co.uk.
1.2SymboleindieserGebrauchsanweisung
WarnungensindindieserGebrauchsanweisungmitSymbolen
gekennzeichnet.DieWarnsymbolesindnebeneiner
Überschriftabgebildet,dieaufdieGefährlichkeithinweist.
WARNUNG
WeistaufeinepotenziellgefährlicheSituation
hin,diebeiNichtbeachtungzumTododerzu
schwerenVerletzungenführenkönnte.
ACHTUNG
Weistaufeinepotentiellgefährliche
Situationhin,diebeiNichtbeachtungzu
Produktbeschädigungenund/oderleichten
Verletzungenführenkönnte.
WICHTIG
WeistaufeinegefährlicheSituationhin,die
beiNichtbeachtungderHinweisezueiner
Produktbeschädigungführenkönnte.
NützlicheTipps,Empfehlungenund
Informationenfüreineefzienteund
reibungsloseVerwendung.
DiesesProduktentsprichtderRichtlinie
93/42/EWGfürMedizinprodukte.Der
MarkteinführungszeitpunktfürdiesesProdukt
istinderCE-Konformitätserklärungfestgelegt.
Hersteller
18PALIN-010-C
Allgemein
1.3Konformität
QualitätistfürdasUnternehmenentscheidend,alleAbläufe
sindnachdenAnforderungenvonISO13485ausgerichtet.
AlleInvacareSoftformMatratzentrageninÜbereinstimmung
mitderRichtlinie93/42/EWGüberMedizinprodukte(Klasse
1)dieCE-Kennzeichnung.
Invacaresetztsichkontinuierlichdafürein,dieUmweltdurch
seineUnternehmenstätigkeitsowohldirektvorOrtalsauch
weltweitmöglichstwenigzubeeinträchtigen.Wirverwenden
ausschließlichREACH-konformeMaterialienundBauteile.
WeitereInformationenerhaltenSiebeiderNiederlassung
vonInvacareinIhremLand(AnschriftenndenSieaufder
RückseitedieserGebrauchsanweisung).
1.4Garantie
WirgewährenfürdasProdukteineHerstellergarantiegemäß
unserenAllgemeinenGeschäftsbedingungenfürdasjeweilige
Land.GarantieansprüchekönnennurüberdenHändler
geltendgemachtwerden,vondemdasProduktbezogen
wurde.
1.5BeschränkungderHaftung
InvacareübernimmtkeineHaftungfürSchädenaufgrund
von:
NichteinhaltungderGebrauchsanweisung
falscherVerwendung
normalemVerschleiß
falscherMontageoderEinrichtungdurchdenKäufer
odereinenDritten
technischenÄnderungen
unbefugtenÄnderungenbzw.Einsatznichtgeeigneter
Ersatzteile
1.6Verwendungszweck
DieseDruckverteilungsmatratzeistzurVerwendung
miteinemBettrahmenentsprechenderGrößeundin
VerbindungmiteinemGesamtpegeprogrammfürdie
Dekubitus-Prophylaxebestimmt.
EineVerwendungzuanderenZweckenistnichtzulässig.
DiesesProduktwurdeimHinblickaufeinewirksame
DruckverteilungfürPatientenbeinormalemGebrauch
entwickelt.Invacarebetrachtetesalseinennormalen
Gebrauch,wenndieLiegeächemiteinemBettlakenaus
Baumwolle,BaumwollmischgewebeoderLeinenbezogen
istundzwischenderLiegeächeunddemBenutzernichts
anderesalseinsolchesLakenliegt.
1.7Produktlebensdauer
DieNutzungsdauerdieserProduktebeträgtschätzungsweise
fünfJahre,vorausgesetzt,siewerdenstrengin
ÜbereinstimmungmitdemindiesemHandbuch
beschriebenenEinsatzzweckverwendetundalleWartungs-
undServiceanforderungenwerdenerfüllt.Beisorgfältigem
UmgangundordnungsgemäßerPegesowieunterder
Voraussetzung,dasstechnischeundwissenschaftliche
FortschrittenichtzutechnischenEinschränkungenführen,
kanndasProduktlängergenutztwerden.Durchhohe
BeanspruchungoderfalschenUmgangoderfalsche
WiederaufbereitungkannsichdieNutzungsdauerauch
reduzieren.DieTatsache,dasswirfürdieseProdukteeine
PALIN-010-C19
InvacarSoftform®
erwarteteLebensdauerangeben,begründetkeinezusätzliche
Garantie.